9. Installing Software
Smart Plus (PHT-35LHS) Installation Manual
133
En
glis
h
Disabling the Packing Mode
Smart Plus
has a unique feature─packing mode
built into
the system to prevent the unit from being damaged while
shipping and transporting. Since it is in the packing mode
through the factory default, the packing mode is required to
be disabled at this step for a successful installation.
If the packing mode is not disabled, the equipment will not
be operated even after being turned on.
1. Click on the
General
tab on the Control Panel.
2. In the
Networking option
, enable the
Serial
checkbox and set
Serial port
and
Baud Rate
as below, and then click
Manager
.
If an error has occurred during connection, make sure the
COM port setting is correct as follows:
1. Run the Device Manager.
2. Check the COM port assigned to “com0com” as shown in
the following figure.
3. If it is not assigned correctly, change the COM port in the
Networking option
and click
Save
.
3. Enter the command
PVER
] to verify the current mode. Note that the equipment is
now in packing mode.