PaX-i
(PCH-2500) Installation Manual
22
2 Choosing an Installation Site
Minimum space required:
●
With Cephalometric unit:
2,820 mm(L)
x 2,147 mm(W) x 2,320 mm(H) / 111”(L) x
85”(W) x 92”(H)
●
Without Cephalometric unit:
1,890 mm x
2,147 mm x 2,320 mm(H) / 75”(L) x 85”(W) x
92”(H)
The system is normally installed beside a wall,
and the operator uses the system on the left.
Lead∙thickness:∙≥1∙mm
The width of the entrance:
The door of the X-Ray room should have a
clearance of more than 800 mm (31.5”) wide.
Floor area:
The floor of the X-Ray room must be stable and
level for system balance.
The floor must be able to support a minimum
weight of 500 kg/m
2
(110 lbs./feet
2
).
Protection against radiation
●
To protect against radiation hazards, follow all
federal and municipal requirements.
●
During exposure, the operator should follow
applicable radiation shielding requirements
and remain at least 2m (7’) from the source of
the radiation.
●
Maintain visible contact with the patient and
a clear view of indicators such as the warning
lamp and imaging status on the PC.
Summary of Contents for Pax-i PCH-2500
Page 1: ...Installation Manual Full version Model PCH 2500 Version 3 06 English...
Page 2: ......
Page 3: ......
Page 46: ...This page is intentionally left blank...
Page 82: ...This page is intentionally left blank...
Page 104: ...This page is intentionally left blank...
Page 124: ...This page is intentionally left blank...
Page 166: ...This page is intentionally left blank...
Page 167: ...Acquiring the Test Image 10...
Page 169: ...11 Technical Specifications...
Page 176: ...This page is intentionally left blank...
Page 179: ...PaX i PCH 2500 Installation Manual 179 Appendix Schematic diagram Components supplied...
Page 183: ...PaX i PCH 2500 Installation Manual 183 Appendix 2 3M 2 2M 2 1M 2 0M 4 Separate the side cover...
Page 199: ......
Page 200: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...