6. Emergency Measures
30
EzRay Premium User Manual
Error Code
Cause
Actions to be taken
E13
The kV feedback is higher
than the set value by more
than + 20 kV during X-ray
irradiation.
"
E14
The mA feedback is lower
than the set value by more
than 50 % during X-ray
irradiation.
"
E15
The mA feedback is higher
than the set value by more
than 50 % during X-ray
irradiation.
"
E60
The exposure switch is
already pressed when you
turned on the equipment.
Check if the exposure switch is
already pressed when you turned on
the equipment.
Once the switch is released, the
equipment functions properly.
If the exposure switch is released but
the equipment does not function
properly, contact the service center.
E61
The exposure switch is
released earlier than the
time specified on the LCD.
The error is reset after 10 seconds.
After the reset, you can perform the
irradiation.
If the error continues, contact the
service center.
Summary of Contents for EzRay Premium VEX-S100W
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Page 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Page 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Page 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Page 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Page 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Page 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Page 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Page 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Page 64: ......