12. X-Ray Dose Data
48
EzRay Premium User Manual
12.2
Leakage Dose Test Report
Test Equipment Information
Instrument
Manufacturer
Model
S / N
Dose Meter
Victorian
660
101114 / 101377
Test Condition
Distance from focal point [m]
1
Applied Tube Voltage Peak [kVp]
70
Applied Tube Current [mA]
6.0
Applied Exposure time [s]
0.5
Direction [
°
]
Basic Cone [mGy/hr]
Transverse
Longitudinal
0
0.0018
0.0605
45
0.0140
0.0395
90
0.0482
0.0535
135
0.0105
0.0561
180
0.0035
0.0947
225
0.0105
0.0561
270
0.0544
0.0614
315
0.0702
0.0509