11. Technical Specifications
EzRay Premium User Manual
43
EN
G
L
ISH
11.2
Technical Specifications
X-Ray Generator
Item
Description
Model
DG-10A05T3
Generator size
Mono-tank: 140 x150 x 70
Weight
2.72 kg (Inverter: 0.92 / Monotask:
1.8)
Rated output power
Max. 0.5 kW
High Voltage
Generator
Tube Voltage (kV)
50 - 70
Tube Current (mA)
4.0 – 7.0
sec
0.04 ~ 2
kV Accuracy
5 %
mA Accuracy
5 %
Cooling
Automatically controlled / Protect ≥
50
℃
Cooling Fan: Optional
Inherent Filtration
At least 1.0 mm Al
Added Filtration
1.0 mm Al
Total Filtration
Min 2.0 mm Al
X-ray Tube
Manufacturer
Toshiba
Model
D-041SB (Stationary Anode Type)
Focal spot size
0.4 mm (IEC60336)
Target Angle
12.5 °
X-ray Coverage
70 mm at SID 200 mm
Duty Cycle
1:60 or more
(Exposure time: interval time)
Summary of Contents for EzRay Premium VEX-S100W
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Page 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Page 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Page 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Page 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Page 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Page 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Page 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Page 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Page 64: ......