5. Using the System
EzRay Premium User Manual
27
EN
G
L
ISH
5.5
X-Ray Generation Limit
Tube Voltage, Tube Current, and Exposure Time value cannot be set
to the maximum value simultaneously.
If the parameter you set exceeds the safety limit, the “Over Value”
message is displayed on the top left of the LCD. Unless this message
is cleared, you cannot perform irradiation.
Tube Voltage (kV)
Tube Current (mA)
Exposure Time (sec)
60 ~ 65
4 ~ 6
0.04 ~ 2.0
61 ~ 70
7
Not available
66 ~ 70
4 ~ 5
0.04 ~ 2.0
60
7
0.04 ~ 0.5
66
6
0.04 ~ 1.8
67
6
0.04 ~ 1.5
68
6
0.04 ~ 1.2
69
6
0.04 ~ 0.9
70
6
0.04 ~ 0.5
Adjustable step: 1 kV, 1 mA, 0.01 sec