Do not use force to unplug cables.
The film holder can be autoclaved up to 145
˚
C or cleaned with alcohol-based
solutions.
Do not keep the equipment or its parts in a humid place or near a liquid
substance.
Avoid placing the equipment in an area that has some degree of slope, vibration,
or shock.
Avoid placing the equipment near chemical storage and gas-filled storage
facilities.
Maintenance Task Checklist
Tasks
Period
Sterilize all components that meet the patient and operator with
an alcohol-based solution.
Daily
Wipe the outer covers of the equipment with a dry cloth at the
end of each day’s operation.
Do not use detergents or solvents to clean the
outer covers of the equipment.
Daily
Ensure sure that the main power switch has been turned off
after using the equipment.
Daily
Ensure that the equipment is firmly plugged into a dedicated
power source.
Daily
Ensure there are no oil leaks.
Daily
Ensure the wall framework is securely attached to the wall.
Daily
Ensure that the plug and power cord are not hot.
Daily
Ensure the power supply cable is in good condition.
Daily
Ensure the audible signal is audible and the x-ray emission
light is visible when you make an exposure.
Daily
Ensure that the orange (exposure) indicator light turns on when
the exposure switch is pressed.
Daily
Ensure that the power cable is not kinked, broken, exposed
and that it is free of all other defects.
Weekly
Ensure that all visible labels are intact and legible.
Monthly
Ensure that the equipment is well-grounded.
Monthly
Summary of Contents for EzRay Premium VEX-S100W
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice vi EzRay Premium User Manual Left blank intentionally...
Page 14: ...1 General and Regulatory Information 6 EzRay Premium User Manual Left blank intentionally...
Page 24: ...3 The System Overview 16 EzRay Premium User Manual Left blank intentionally...
Page 36: ...5 Using the System 28 EzRay Premium User Manual Left blank intentionally...
Page 40: ...6 Emergency Measures 32 EzRay Premium User Manual Left blank intentionally...
Page 46: ...9 Service 38 EzRay Premium User Manual Left blank intentionally...
Page 48: ...10 Disposing of the Unit 40 EzRay Premium User Manual Left blank intentionally...
Page 52: ...11 Technical Specifications 44 EzRay Premium User Manual...
Page 58: ...12 X Ray Dose Data 50 EzRay Premium User Manual Measurement Values Longitudinal...
Page 64: ......