ProStar 335 Photodiode Array Detector Control Manual - 35 -
Peak Sensor Control
Peak Sense Relay group:
In the Peak Sense Relay Active scrolling list, select Open or
Closed. If your selection is Close on activation, the relay is
initially open. Conversely, if you command Open on activation,
the relay will be initially closed.
Enter the Duration time in the corresponding field. The Duration
is the length of time the relay is pulsed, whether closed, or open.
The time entered for the Duration will be dictated by the
specifications of the particular sample collector or column
switching valve used. Generally a duration of 0.5 seconds will
satisfy most applications.