ProStar 335 Photodiode Array Detector Control Manual - 18 -
Option Window
Select the Cell Type according to the one physically installed in
the detector, 4 cell types are selectable the 9x0, 9x1, 4x0 and
4x0.15.
Check or uncheck the Normalize check box to activate or not
this option. This option is active only with a 9x0 flow cell and
scales the absorbance results to a 10 mm flow cell to allow
absorbance comparison. The scale factor applied to the
absorbance is 10/9.
If a prep flow cell is selected, the user will be able to enter a
Cell Ratio. The ratio is used for extended range operation. The
ratio for each flow cell is determined at the factory and is
included with the flow cell. The detector will always operate in
extended range mode with a prep flow cell.
Summary of Contents for ProStar 335