Models 947, 948 and 960 Leak Detectors Operations and Service Manual
A-8
[SHIFT]
SET
STD LK
SHOW
Sets the leak rate value to which the leak detector calibrates. Enter
X.XE-X (the desired value), followed by the SHIFT key, then the
function key.
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #1. Enter X.XE-X
(the desired value), then the function key.
[SHIFT]
REJ3
SET
REJ 1
Sets the leak rate value for Reject Set Point #3. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #2. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.
[SHIFT]
REJ 4
SET
REJ 2
Sets the leak rate value for Reject Set Point #4. Enter X.XE-X
(the desired value), followed by the SHIFT key, then the
function key.