#6/27401
46
Varian ELSD Service Manual
Replacing The Varian 385-LC Cooled Evaporator Assembly
When required:
Failure or damage to the Evaporator Heater Assembly / Evaporation tube
become irreversibly contaminated
Tools required:
Allen keys
Spanner
Parts required:
Varian 385-LC Peltier Assembly
(#PL0890-0630
)
If this procedure is not followed correctly, the cooled evaporator
assembly may be damaged
Open the detector’s side panel using
the access key, as shown.
Remove the evaporator tube from the
nebuliser chamber as described in
“
Replacing the nebuliser chamber”
(pg
24)
Do not discard evaporator cartridge.