Varian ELSD Service Manual
Disconnect the two
Evaporator Heater
Assembly cables from the
main control board
(EVAP.HTR & EVAP) and
remove the evaporator
heater assembly.
Insert the New Evaporator
Heater Assembly into the
optical block.
Insert the Cartridge into the
Evaporator heater assembly
and insert assembly into the
nebuliser chamber.
Connect cables (EVAP.HTR
& EVAP) to main control
board
Ensure the evaporator cartridge is inserted into the evaporator
assembly prior to reassembly; otherwise the instrument will not
perform correctly