#6/27401
42
Varian ELSD Service Manual
Remove the light trap, to allow easier
access to the regulator.
See section
“Replacing the Optical light
trap”
(pg 59 ) for details of how to remove
the light trap
In order to remove the regulator, undo
these two screws.
Detach all gas lines into the regulator and
extract the regulator
Insert the new regulator, re-fasten the
screws, reconnect the gas tubing and re-
attach the light trap
Follow the procedure outlined in “
Installing
the nebuliser” (pg 20)
in order to configure
the correct gas flow to the nebuliser