Varian ELSD Service Manual
Disconnect all 3 cables from the PMT
assembly.
Loosen the grub screw on the PMT assembly
and remove the black rubber plug, as shown.
Carefully remove the PMT assembly from the
optical block.
Attach the replacement PMT assembly
making sure all three cables are re-
connected.
Tighten the Grub screw and re-insert the
black rubber plug.
The rubber plug must be inserted into the PMT assembly,
otherwise the detector will not perform correctly.