Varian ELSD Service Manual
Replacing The Evaporator Heater Assembly (Evaporation Tube)
When required:
Faulty Heater / Evaporation tube become irreversibly contaminated
Tools required:
Allen keys
Spanner
Parts required:
Evaporator Heater Assembly (No Cartridge) (# PL0890-0480)
Open the detector’s side
panel using the access key,
as shown.
Remove the evaporator tube
from the nebuliser chamber
as described in “
Replacing
the nebuliser chamber”
(pg
24)
Do not discard evaporator
cartridge.
Carefully remove the
opposite end of the
evaporator tube from the
optical block assembly.