10. H OT A IR G EN ER ATO R
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – Guidance
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Contact ±8 kV
Air ±15 kV
5/50 ns,
100 kHz, ±2 kV
5/50 ns,
100 kHz, ±2 kV
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be similar to
that of a typical residen-
tial, commercial or
hospital environment. If
the user requires
continuous function, it is
recommended that the
VOLCANO MEDIC 2
Vaporization System is
operated from an
uninterruptible power
supply or battery.
Contact ±8 kV
Air ±15 kV