10. HOT A IR G EN ER ATO R
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – Guidance
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
LtL: ± 1 kV
LtG: ± 2 kV
0 % Uт
for 0.5 cycles (1
phase) at 0, 45,
90, 135, 180,
225, 270 and
315 angle
degrees
0 % Uт for 1
cycle and
70 % Uт for
25/30 cycles
(50/60 Hz)
single-phase at
0 angle degrees
0 % Uт for
250/300 cycles
(50/60 Hz)
LtL: ± 1 kV
LtG: ± 2 kV
0 % Uт
for 0.5 cycles (1
phase) at 0, 45,
90, 135, 180,
225, 270 and
315 angle
degrees
0 % Uт for 1
cycle and
70 % Uт for
25/30 cycles
(50/60 Hz)
single-phase at
0 angle degrees
0 % Uт for
250/300 cycles
(50/60 Hz)
Mains power quality
should be similar to
that of a typical residen-
tial, commercial or
hospital environment. If
the user requires
continuous function, it is
recommended that the
VOLCANO MEDIC 2
Vaporization System is
operated from an
uninterruptible power
supply or battery.
Mains power quality
should be similar to that
of a typical residential,
commercial or hospital
environment. If the user
requires continuous
function, it is recom-
mended that the
VOLCANO MEDIC 2
Vaporization System is
operated from an
uninterruptible power
supply or battery.