Performance Characteristics
T
e
chnical S
p
ecifications
SurgiStat II-20 User’s Guide
8
-
3
Operating Parameters
Transport and Storage
Audio Volume
The audio levels stated below are for activation tones (Bipolar, Cut, and Coag)
and alarm tones (return electrode and system alarms) at a distance of one meter.
Alarm tones meet the requirements for IEC 60601-2-2.
Activation Tone
Ambient temperature
range
10° to 40°
C (50° to 104° F)
Relative humidity
15% to 75%, noncondensing
Atmospheric pressure
700 to 1060 millibars
Warm-up time
If transported or stored at temperatures outside the
operating temperature range, allow one hour for the
generator to reach room temperature before use.
Ambient temperature
range
-34° to 65° C (-29° to 149° F)
Relative humidity
0% to 75%, noncondensing
Atmospheric pressure
500 to 1060 millibars
Volume (adjustable)
45 to 65 dBA
Frequency
Cut: 1 kHz
Blend: 1 kHz
Desiccation: 2 kHz
Fulguration: 2 kHz
Bipolar: 2 kHz
Duration
Continuous while the generator is activated