Initial Inspection
3-2
SurgiStat II-20 User’s Guide
Initial Inspection
When you first unpack your SurgiStat II generator, inspect it visually:
•
Look for any signs of damage.
•
Verify that the shipping package contains all items listed on the packing list.
If the unit or any accessories are damaged, notify Valleylab Customer Service
immediately. Do not use any damaged equipment.
Installation
Place the SurgiStat II on any flat surface with a tilt angle not more than 10°. The
unit relies on natural convection cooling. Do not block its bottom or rear vents.
Ensure that air flows freely on all sides of the unit.
Function Checks
Upon initial installation of the unit, perform the following checks. Refer to the
figures in the previous chapter for the location of connectors and controls.
Setting Up the Unit
1.
Verify that the Power Switch is in the Off (
O
) position and that no accessories
are connected to the unit.
2.
Connect a hospital grade power cable to the AC power cable receptacle on the
back of the unit, then to a properly grounded wall outlet.
3.
Connect a two-pedal footswitch to the appropriate footswitch receptacle on
the back of the unit. Use only a Valleylab footswitch with the SurgiStat II-20
generator. Although other types of footswitches may fit, they may not be
compatible. Use of an incompatible footswitch may cause unexpected output.
Warning
Connect the power cord to a properly polarized and grounded power source with
the frequency and voltage characteristics that match those listed on the back of
the unit.
Caution
At no time should you touch the active electrode or bipolar forceps. A burn could
result.
Summary of Contents for SurgiStat II-20
Page 1: ...User s Guide SurgiStat II 20 Electrosurgical Generator...
Page 14: ...1 8 SurgiStat II 20 User s Guide...
Page 24: ...2 10 SurgiStat II 20 User s Guide...
Page 38: ...4 10 SurgiStat II 20 User s Guide...
Page 42: ...5 4 SurgiStat II 20 User s Guide...
Page 46: ...6 4 SurgiStat II 20 User s Guide...
Page 50: ...7 4 SurgiStat II 20 User s Guide...