Output Characteristics
T
e
chnical S
p
ecifications
SurgiStat II-20 User’s Guide
8
-
7
Output Characteristics
Maximum Output for Bipolar and Monopolar Modes
Power readouts agree with actual power into rated load to within 20% or 5 watts,
whichever is greater. All measurements were taken at the nominal input power
voltage used for calibration.
* An indication of a waveform’s ability to coagulate bleeders without a cutting
effect
Mode
Output Power
Output
Frequency
Repetition Rate
Vp-p max
Crest Factor*
(Rated Load)
Cut
120 W @ 500
Ω
357 kHz ± 50
kHz
N/A
2.5 kV
2.9 ± 20%
Blend
90 W @ 800
Ω
357 kHz ± 50
kHz
30 kHz ± 5 kHz
3.5 kV
3.3 ± 20%
Desiccate
80 W @ 1000
Ω
475 kHz ± 50
kHz
57 kHz ± 5 kHz
4.5 kV
5.5 ± 20%
Fulgurate
40 W @ 1000
Ω
410 kHz ± 50
kHz
25 kHz ± 5 kHz
6.5 kV
7.7 ± 20%
Bipolar
30 W @ 200
Ω
520 kHz ± 50
kHz
32 kHz ± 5 kHz
2.0 kV
6.9 ± 20%