ForceTriad User’s Guide
9-11
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth
conductor.
Type CF Equipment (IEC 60601-1)/Defibrillator Proof
This generator provides a high degree of protection against electric
shock, particularly regarding allowable leakage currents. It is type CF
isolated (floating) output and may be used for procedures involving the
heart.
This generator complies with the ANSI/AAMI HF18 specifications for
“defibrillator proof” designation and IEC 60601-2-2.
Liquid Spillage (IEC 60601-2-2 Clause 44.3)
The ForceTriad energy platform is constructed so that liquid spillage in normal
use does not wet electrical insulation or other components which when wetted are
likely to adversely affect the safety of the equipment.
Voltage Transients (Emergency Energy Platform Mains Transfer)
The ForceTriad energy platform continues to operate normally with no errors or
system failures when transfer is made between line AC and an emergency energy
platform voltage source. (IEC 60601-2-2 sub-clause 51.101 and AAMI HF18
sub-clause 4.2.2)