After Surgery
Li
gaS
u
re T
iss
ue Fusi
on
ForceTriad User’s Guide
6-9
Reactivate
A
four-pulsed
tone will sound when the Reactivate screen is displayed.
If this message appears the user should:
1.
Release the footswitch pedal or hand switching button.
2.
Reactivate the seal cycle without repositioning the instrument.
Endpoint not reached
– Additional time and energy are needed to complete the
fusion cycle.
Seal cycle interrupted
– The seal cycle was interrupted before completion. The
handswitch or footswitch was released before the end tone activated.
Reactivate / Energize
– Reactivate the seal cycle without removing or
repositioning the instrument.
After Surgery
Disconnect the instruments
1.
Turn off the energy platform.
2.
Disconnect all instruments from the front panel.
•
If the instrument is single use only (disposable), dispose of it according to
the procedures for your institution.
•
If the instrument is reusable, clean and sterilize it according to the
manufacturer’s instructions for use.
3.
Disconnect and store any footswitch(es) used.