Performance Characteristics
9-8
ForceTriad User’s Guide
Power Cord Specification
This unit was equipped from the factory with a 110 VAC hospital grade NEMA
5-15 power cord. Should the AC power cord need to be replaced to match another
plug configuration, the replacement plug/cable/receptacle configuration must
meet or exceed the following specifications:
100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft (5 m)
Plug - minimum 10 A - 125 VAC
Unit receptacle - IEC female, minimum 10 A - 125 VAC
220-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15 ft (5 m)
Plug - minimum 6 A - 250 VAC
Unit receptacle - IEC female, minimum 6 A - 250 VAC
Input Frequency
The ForceTriad energy platform operates within specification at all line input
frequencies between 48 Hz and 62 Hz. The User does not need to reconfigure the
ForceTriad energy platform for different line frequencies.
Input Current
The ForceTriad energy platform draws no more than 10A at any line input
voltage.
Backup Power
The ForceTriad energy platform retains all user programmed features, calibration,
and statistical data when switched off and unplugged. The ForceTriad energy
platform operates within specification when switched over to a supplied line
power by hospital backup systems.
Equipotential Ground Connection
An equipotential ground connection is provided to allow connection of the
ForceTriad energy platform to ground.
ECG Blanking
An ECG blanking port is provided to signal other devices that the ForceTriad
energy platform is active. The receptacle is a 2.5 mm mono jack. It is electrically
isolated from the internal ground referenced electronics with the shell electrically
connected to the chassis for ESD protection.