Output Characteristics
Maximum Output for Bipolar, Monopolar, and LigaSure Modes
Power readouts agree with actual power into rated load to within 15% or 5 watts,
whichever is greater.
* An indication of a waveform’s ability to coagulate bleeders without a cutting effect.
Mode
Open Circuit Peak
Voltage (max)
Open Circuit P–P
Voltage (max)
Rated Load
(max)
Power
(max)
Crest
Factor*
Duty
Cycle
Bipolar
Low
Standard
Macro
250 V
175 V
250 V
500 V
350 V
500 V
100
Ω
100
Ω
100
Ω
95 W
95 W
95 W
1.42
1.42
1.42
N/A
N/A
N/A
Monopolar Cut
Cut
Blend
920 V
1485 V
1840 V
2970 V
300
Ω
300
Ω
300 W
200 W
1.42
2.7
N/A
50%
Valleylab (HWD)
2365 V
4730 V
300
Ω
200 W
4.3
25%
Monopolar Coag
Fulgurate
Spray
3050 V
3625 V
6100 V
7250 V
500
Ω
500
Ω
120 W
120 W
5.55
6.6
6.5%
4.6%
LigaSure
287.5 V
575 V
20
Ω
350 W
1.42
N/A