69
V.
When prompted by the display, apply 300mV to the sense
terminals. Press
“E
NTER
”
to continue or
“C
LEAR
”
to return to
the 10mV cal screen. While calibrating the ranges, the
screen will display the following information:
VI.
When prompted by the display, apply 2V to the sense
terminals. Press
“E
NTER
”
to continue or
“C
LEAR
”
to return to
the 300mV cal screen. While calibrating the ranges, the
screen will display the following information:
VII.
When prompted by the display, press
“E
NTER
”
to save the
calibration values to memory, or press
“C
LEAR
”
to return to
the 2V calibration screen.
VIII.
When prompted by the display, press
“E
NTER
”
to exit the
calibration mode, or press TCM to continue with the TCM
Cal.