TCS – Temperature Compensator Setup
As earlier mentioned, the 4300C needs to reference a temperature coefficient and a temperature
reference to calculate the compensated value. Stored in the instrument’s memory are six configurations
that can be selected. The table below illustrates the values associated to each of the configurations.
The following procedure illustrates how to select one of these configurations and how to setup a
customized configuration. This procedure does not need to be repeated every time the TCM mode is
selected. The configuration that is selected or the custom values inputted will be set as default.
Press “
F
UNC
. + TCM + 1”
to bring up the first selection screen.
Press the number associated with one of the configurations, and
press “
E
NTER
”
or go to the next selection screen by pressing
“F
UNC
.”
Press the number associated with one of the configurations, and
press “
E
NTER
”
or go to the next selection screen by pressing
“F
UNC
.”.
(Press
“C
LEAR
”
to return to the previous screen.)
Press
“7 + E
NTER
”
to enter the custom configuration screen. (Press
“
C
LEAR
” to return to the previous screen.)
Using the numeric keys input a value for the temperature
coefficient. (Use
“C
LEAR
”
as a backspace to correct errors.)
Press
“E
NTER
”
to confirm the value.
Using the numeric keys input a value for the reference temperature.
(Use
“C
LEAR
”
as a backspace to correct errors.)
Press
“E
NTER
”
to confirm the value.
The screen will display the values entered.
Material
Configuration
Temperature
Coefficient
Temperature
Reference
Copper
CU20
3931ppmΩ/°C
20°C
CU20
25°C
Aluminum
AL20
4030ppmΩ/°C
20°C
AL25
25°C
Gold
AG20
3000ppmΩ/°C
20°C
AG25
25°C