63
RDNG? QUERY
Responds with reading from the device in engineering notation. Query function also uses
an immediate update of the conversion so repeated query requests gets the most
accurate data.
Syntax:
RDNG?
Example:
RDNG?<crlf>
2.4321e+1<crlf>
(on 20.000Ω ranges)
Response:
<value in reduced engineering notation<crlf>>
Power-on default = 0.000e+0
CAUTION: Will not work if the transmit interrupt happens to be busy with a display update
and the following occurs:
Remote
->RANGE 6
<2kΩ)
System
->Places "REM^" in TX buffer (Tiny Display)
System
->Change to R6
System
->Places "2k^" in TX buffer (Tiny Display)
System
->Formats Display String "1.6543k^"
Remote
->READING?
System
->Places in TX buffer to VFD 'D' characters
Remote
->RANGE 4
System
->Change to R4
System
->Executes READING? query, uses Range=4 and last displayed value
System
->Places in TX buffer, next 'R' characters
RESET COMMAND
Executes a soft reset of the ADuC834 processor system.
Syntax:
RESET
Response:
Front panel display show soft reset initiation. "RESETTING" is displayed
flashing inverse mode for 300ms while all system configurations are
returned to power up default.
Example:
RESET<crlf>
Response:
<crlf>
Notes: Resetcpu() executes as soft reset of the system by "nicely" changing the return
stack function and popping it. Expect a 500ms delay after receiving the linefeed before
transmitting next command.