28
Press
“E
NTER
”
to confirm the values or
“C
LEAR
”
to return to the
previous screen.
The screen will display a brief message and the instrument will
return to the standard measurement mode.
TCC – Temperature Compensator Calibration
This procedure is used to match a 4300C with its Omni Compensator.
Verify that the instrument is in a temperature-controlled area.
Press “
F
UNC
. + TCM + 2”
to initiate the process.
Connect the Omni Compensator to the TCM receptacle on the
front panel of the Ohmmeter.
Allow the Omni Compensator to reach ambient temperature.
(Wait 5 to 10 minute before proceeding.)
Press “
E
NTER
” to continue.
Using the numeric keys input a value for the ambient
temperature. (Use “Clear” as a backspace to correct errors.)
Press “
E
NTER
” to confirm the value.
Press “
E
NTER
” to confirm the value or “
C
LEAR
” to return to the
previous screen.
Once the calibration values are stored, press “
TCM
” and verify
that the temperature displayed matches the ambient
temperature. If the values do not match, repeat the procedure.