Chapter 3 _____________________________________________ Constructing an Ozone Sounding
VAISALA_______________________________________________________________________ 61
Preparing the Radiosonde and
OIF411
Connecting Ozone Sensor Pump to OIF411
-
Cable needed: CBL210282
Connect the cable to the OIF411 terminal marked
Pump
.
Figure 27
Connecting Ozone Sensor Pump Cable
CAUTION
To avoid any problems with the sounding, do not use the sensor loaded
(pump-powered) with solutions, if the sensor is not connected to a
powered interface, or if the anode and cathode wires are connected.
The pump is allowed to be used only with NO-OZONE, or when the
destruction filter is connected. Never use HI-OZONE.