User’s Guide ______________________________________________________________________
100 __________________________________________________________________M211486EN-A
xdata= <Instrument type is 05><Instrument number is 01><Ozone
pump T [0.01 ºC]><Ozone current [0.0001uA]><Battery voltage
[0.1V]> <Ozone pump current [1 mA]><Ext. voltage [0.1V]>CR
This means that if OIF411 sends
xdata=050108CA186A0750B637CR
it is interpreted as shown in
ID Data
The radiosonde also sends OIF411 ID data, which contains:
-
Instrument type and number
-
OIF411 serial number
-
Diagnostics word
-
Software version once per minute
OIF411 ID data replaces OIF411 measurement data, but not the
additional sensor data, sent once per minute.
The OIF411 ID data contains the following information:
xdata= <Instrument type is 05><Instrument number is 01><OIF serial
number><diagnostics word><SW version><I>CR.
This means that if OIF411 sends
xdata=0501G12345670001000AICR
Table 15
OIF411 Data Interpretation 1
Instrument
Type
2 Bytes
Instrument
Number
2 Bytes
Ozone
Pump T
[0.01 ºC]
4 Bytes.
MSB bit is
a sign bit.
Ozone
Current
[0.0001u
A]
5 Bytes
Battery
Voltage
[0.1V]
2 Bytes
Ozone
Pump
Current [1
mA]
3 Bytes
Ext.
Voltage
[0.1 V]
2 Bytes
Hex ASCII 05
01
08CA
186A0
75
0B6
37
Interpreted
as
5
1
22.50
10.0000 11.7
182
5.5