82
PR-23 instruction manual
9.3.3 PR-23-AC specifications
General specifications
Refractive Index range:
Full range n
D
1.3200–1.5300 (corresponds to hot
water – 100 Brix)
Accuracy:
Refractive index n
D
± 0.0002 (corresponds typi-
cally to ± 0.1 % by weight)
Repeatability and stability correspond to accuracy
Speed of response:
1 s undamped, damping time selectable up to 5 min
Calibration:
With Cargille certified refractive index liquids over
full range of n
D
1.3200–1.5300
CORE-Optics:
No mechanical adjustments (US Patent No. US6067151)
Digital measurement:
3648 pixel CCD element
Light source:
Light emitting diode (LED) 589 nm wavelength,
sodium light
Temperature sensor:
Built-in Pt-1000
Temperature compensation:
Automatic, digital compensation
Instrument verification:
With certified refractive index liquids and
Vaisala documented procedure
Ambient temperature:
Sensor: max. 45 °C (113 °F),
min. -20 °C (-4 °F)
Indicating transmitter: max. 50 °C (122 °F),
min. 0 °C (32 °F)
SENSOR PR-23-AC:
Compact sensor model for small pipe lines
Process connection:
Sanitary 3A-clamp 2.5"; Varivent
®
in-line access
unit clamp DN65 or via elbow flowcell (for line
sizes of 2.5" and smaller)
Sanitary design:
3-A Sanitary Standard 46-03 approved and
EHEDG (European Hygienic Engineering & Design
Group) certified (connections -H, -E, -N)
Process pressure:
Sanitary clamp max. 15 bar (200 psi) at 20 °C (70 °F)/9 bar
(125 psi) at 120 °C (250 °F)
Process temperature:
-20 °C–+130 °C (-4 °F–+266 °F)
Process wetted parts, standard:
AISI 316L stainless steel, prism spinel, prism gas-
kets MTF (Modified Teflon)
Sensor protection class:
IP67, Nema 4X
Sensor weight:
2.0 kg (4.4 lbs)
Summary of Contents for K-PATENTS PR-23 Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...