134
PR-23 instruction manual
9.10 PR-23 process refractometers in potentially
explosive atmosphere
The PR-23 refractometer series can be used in locations with potentially explosive
atmosphere
with following modifications, made by Vaisala Oyj
. The refractometer sen-
sor’s compliance with the Essential Health and Safety Requirements is assured by
complying with standard EN 50 021:1999.
The PR-23-...-AX refractometers have been certified by DEKRA Certification B.V. un-
der the European ATEX directive 2014/34/EU for ATEX Ex II 3G / Ex nA IIC T4 Gc
and under the IECEx scheme for Ex nA IIC T4 Gc. The Type-examination Certifi-
cate number is KEMA 05ATEX1183X and the IECEx Certificate number is IECEx KEM
10.0033X. These certifications cover the following Ex standards: EN 60079-0:2012 /
IEC 60079-0:2011, EN 60079-15:2010 / IEC 60079-15:2010.
The PR-23-...-FM refractometers are certified by Factory Mutual Research Corpora-
tion, Approval ID 3026104. Equipment Ratings: Nonincendive for use in Class I, Divi-
sion 2, Groups A, B, C & D, Hazardous (Classified) Locations. Temperature identifica-
tion rating for PR-23-...-FM is T6 (T
amb
= 45 °C).
The PR-23-...-CS refractometers are certified by Canadian Standards Association for
Class I, Division 2, Groups A-D. The Certificate Number is 1706327. Equipment Rat-
ings: Nonincendive for use in Class I, Division 2, Groups A, B, C & D, Hazardous (Clas-
sified) Locations. Temperature identification rating for PR-23-...-CS is T4 (T
amb
= 45
°C).
9.10.1 Equipment
The refractometer system (Figure 9.31) for potentially explosive atmosphere loca-
tions consists of a modified refractometer sensor PR-23-...-AX/FM/CS, a standard In-
dicating transmitter DTR and a sensor cable PR-8230-...
Hazardous area
(up to Zone 2)
Safe Area
PROCESS INSTRUMENTS
POWER
WIRING SEE WRG-342
Figure 9.31
Refractometer system PR-23-...-AX/FM/CS
Summary of Contents for K-PATENTS PR-23 Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...