7 Regular maintenance
61
7 Regular maintenance
The need for regular maintenance is minimal, due to the construction with no moving
parts, no mechanical adjustementsand with a solid-state light source. The following
rules apply:
–
Keep the sensor head and the Indicating transmitter clean and dry.
–
Check that the ambient temperature is not above +45 °C (113 °F). The sensor head
should not be too hot to keep a hand on.
–
If your refractometer has prism wash, check that it works, see Section 5.1.3.
–
Follow preventive maintenance program, if any.
7.1 Checking the sensor humidity level
The PR-23 sensor head has an internal humidity detector. The humidity reading can
be checked on the Indicating transmitter display, select
3 SENSOR STATUS
from the Main
menu.
Check the humidity reading once in every three months.
Increasing humidity level indicates either condensate forming in the sensor head (if
the process temperature is below ambient) or prism leakage. If the humidity reading
exceeds 30 %, replace the dryer. If the reading exceeds 50 %, check the prism seals.
Relative humidity exceeding 60 % will produce a diagnostic message
HIGH SENSOR
HUMIDITY
(see Section 8.1.7). Contact service if internal humidity increases.
7.2 Checking the prism and prism gaskets
Once a year
check that the prism surface is smooth and clean and free of erosion and
small holes or digs. If the prism is scratched, eroded, or the gaskets seem to leak,
contact service.
Important:
In
3-A certified sensors
, prism gasket replacement and other repairs
must be done
by authorized Vaisala K-PATENTS
®
service centers only
. If gaskets
of 3-A certified sensors are replaced in the field, the certification is no more valid.
Summary of Contents for K-PATENTS PR-23 Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...