138
PR-23 instruction manual
9.11 Intrinsically safe refractometers PR-23-...-IA and PR-23-...-IF
Hazardous locations are places where a possibility of fire or explosion exists because
of flammable gases, vapors or fine dust.
Zone 0:
An area in which an explosive gas-air mixture is continuously present
or present for long periods of time.
Zone 1:
An area in which an explosive gas-air mixture is likely to occur in nor-
mal operation.
Intrinsically Safe Process Refractometer PR-23-...-IA/-IF can be used in hazardous lo-
cations in Zone 0 and Zone 1 areas.
The PR-23-...-IA refractometers have been certified by Eurofins Expert Services Oy un-
der the European ATEX directive 2014/34/EU
for ATEX Ex II 1G and Ex I M1 / Ex ia
IIC T4 Ga and Ex ia I Ma (T
amb
= -20 – + 65°C)
and under the IECEx scheme
for Ex
ia IIC T4 Ga and Ex ia I Ma (T
amb
= -20 – +65°C)
. The EU Type-examination Certifi-
cate number is EESF 19 ATEX 028X and the IECEx Certificate number is IECEx EESF
19.0010X. These certifications cover the following Ex standards: EN 60079-0:2012 /
IEC 60079-0:2011 and EN 60079-11:2012 / IEC 60079-11:2011.
The PR-23-...-IF refractometer is certified by FM under the United States standards
for IS/I/1/ABCD/T4 and I/0/AEx ia/IIC/T4 (T
amb
= -20 – +45 °C)
The certificate
ID Number is 3036400. This certification covers the following US standards: class
3600 1998, class 3610:2007, class 3810:2005, ANSI/ISA-12.00.01:1999, ANSI/ISA-
12.02.01:2002, ANSI/ISA-82.02.01:2004, ANSI/NEMA 250:1991 and ANSI/IEC
60529:2004.
The PR-23-...-IF refractometer is certified by FM under the Canadian standards
for
IS/I/1/ABCD/T4 and I/0/Ex ia/IIC/T4 (T
amb
= -20 – +45 °C)
. The certificate ID
Number is 3036400C. This certification covers the following Canadian standards:
CSA C22.2 No. 94:1999, CSA C22.2 No. 142:2004, CSA C22.2 No. 157:2006, CSA
C22.2 No. 60529:2005, CSA C22.2 No. 61010.1-1:2004, CSA E60079-0:2007 and CSA
E60079-11:2002.
Note:
Servicing of the intrinsically safe PR-23-...-IA/-IF refractometer is only allowed
for trained service personnel of Vaisala and its representatives. Servicing must be
done according to separate instructions defined by Vaisala and must be reported to
Vaisala.
9.11.1 Equipment
Intrinsically Safe Process Refractometer consists of: a modified refractometer sen-
sor PR-23-...-IA/-IF, an Indicating transmitter STR with single sensor connectivity, and
IS Isolator and cabling between the refractometer sensor and the transmitter (Fig-
ure 9.34).
The equipment is intrinsically safe only if
all
mounting instructions in Section 9.11.2
are followed. If the instrument has been in any way damaged during transportation,
return it to your nearest Vaisala service point for checkup before installation. Never
install a damaged instrument into the process line.
Summary of Contents for K-PATENTS PR-23 Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...