6
PR-23 instruction manual
2.2 Mounting the sensor
The sensor mounting location should be chosen with care to ensure reliable read-
ings from the process. Some basic rules, described in this section, apply to all sensor
models. The model specific instructions can be found in Chapter 9, “Sensor specifica-
tions”. For the Sanitary compact refractometer PR-23-AC see Section 9.3, for the Probe
sanitary refractometer PR-23-AP see Section 9.4, for the Process probe refractometer
PR-23-GP see Section 9.6, for the Teflon body refractometer PR-23-M and Teflon body
semicon refractometer PR-23-MS see Section 9.8 and for the Saunders body refrac-
tometer PR-23-W see Section 9.9. For mounting an ATEX/FM/CSA approved sensor
in explosive atmosphere, see Section 9.10.2. For mounting an intrinsically safe refrac-
tometer PR-23-...-IA, see Section 9.11. For mounting of the Safe-Drive™ system with
the PR-23-SD sensor, see Section 11.4.
2.2.1 Choosing sensor mounting location
A PR-23 refractometer sensor can be located either indoors or outdoors in most cli-
mates. However, when a sensor is located outdoors, some basic protection against
direct exposure to sunlight and rain should be provided. Special care should be taken
if the pipe wall is translucent (e.g. of fiberglass), as light from outside reaching the
prism through the pipe wall may disturb the measurement.
The mounting location needs to be such that sediments or gas bubbles cannot accu-
mulate by the sensor. Good flow velocity is essential in keeping the prism clean.
Important:
If the process pipe vibrates, support the pipe. A vibrating pipe might
damage the in-line sensor mounted on it.
Always check that the sensor head is kept cool enough; the sensor head should not be
too hot to keep a hand on. The sensor cover should not be exposed to high temperature
radiation. In most cases, draft and natural convection provide sufficient air cooling if
the air gets to flow freely around the sensor head.
Additional cooling is necessary when the ambient temperature is higher than 45 °C
(113 °F) or when the process temperature is above 110 °C (230 °F) and the ambient
temperature is above 35 °C (95 °F). The air cooling is improved by blowing pressurized
air against the sensor cover. The pressurized air can be supplied by the ventilation
system. If no air is available it is also possible to install water cooling with PR-10038
cooling cover (except for PR-23-SD where sensor head needs to be kept in original
size for insertion and retraction).
Important:
Always mount the sensor so that the interconnecting cable points down-
wards from the sensor head.
Summary of Contents for K-PATENTS PR-23 Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...