A.4 Modbus registers
Registers are numbered in decimal, starting from 1. Register addresses in actual Modbus
messages (Modbus Protocol Data Unit (PDU)) are in hexadecimal and start from zero. Register
number 1 corresponds to address 0
hex
in the actual Modbus message.
Reading the wrong register(s) may result in correct-looking values.
Check the reference documentation of your Modbus host (PLC) to verify which
notation it uses for Modbus register addresses.
CAUTION!
A.4.1 Measurement data registers
Table 75 Floating point measurement data registers (read-only)
Register
number
Address
Register description
Data format
Unit
1
0000
hex
Relative humidity
32-bit float
%RH
0001
hex
3
0002
hex
Temperature
32-bit float
°C
0003
hex
7
0006
hex
Dew point temperature
32-bit float
°C
0007
hex
9
0008
hex
Dew/frost point temperature 32-bit float
°C
0009
hex
11
000A
hex
Dew/frost point temperature
at 1 atm
32-bit float
°C
000B
hex
13
000C
hex
Dew point temperature at 1
atm
32-bit float
°C
000D
hex
15
000E
hex
Absolute humidity
32-bit float
g/m
3
000F
hex
17
0010
hex
Mixing ratio
32-bit float
g/kg
0011
hex
19
0012
hex
Wet-bulb temperature
32-bit float
°C
0013
hex
Appendix A – Modbus reference