1. Use a 4-mm Allen key to open the hex screws and remove the top part of the probe
holder.
2. Hold the base part against the mounting surface and mark the locations of the 2 screw
holes.
3. Drill holes at the marked locations using a 6-mm drill bit. Make the holes at least 30 mm
deep.
4. Insert the wall plugs in the holes.
5. Mount the base part using screws.
6. Insert the probe body in the holder.
7. Place the top part on the base and tighten the hex screws with a 4-mm Allen key.
4.2 Wiring
1
5
3
4
2
Figure 4 M12 5-pin A-coded male connector pinout
Pin #
Function
Notes
Wire colors in Vaisala cables
1
Power supply
Operating voltage:
• HMP7: 18 ... 30 V DC
• Other models: 15 ... 30 V DC
Current consumption: 10 mA typical,
500 mA max.
Brown
2
RS-485 -
White
3
Power GND and
RS-485 common
Blue
4
RS-485 +
Black
5
Not connected
Gray