0
0.1
0.2
0.3
0.4
A
cc
ur
ac
y
[±
°
C]
Temperature [°C]
-4
0
-3
0
-2
0
-1
0
0
10
20
3
0
4
0
5
0
6
0
0.5
Figure 34 HMP1 temperature measurement accuracy over full range
Table 7 HMP1 operating environment
Property
Description/Value
Operating temperature
−40 … +60 °C (−40 … +140 °F)
Storage temperature
−40 … +60 °C (−40 … +140 °F)
Measurement environment
For air, nitrogen, hydrogen, argon, helium, and
oxygen
IP rating
IP50: Limited protection against dust
1)
Consult Vaisala if other chemicals are present. Consider safety regulations with flammable
gases.
Table 8 HMP1 inputs and outputs
Property
Description/Value
Operating voltage
15 … 30 V DC
Current consumption
2 mA typical, 200 mA max.
Digital output
RS-485, non-isolated
Protocol
Modbus RTU
Table 9 HMP1 output parameters
Output parameter name and unit
Absolute humidity (g/m
3
)
Relative humidity (%RH)
Absolute humidity at NTP (g/m
3
)
Relative humidity (dew/frost) (%RH)
Dew point temperature (°C)
Temperature (°C)
Chapter 9 – Technical data
65
Summary of Contents for HMP Series
Page 10: ...HMP Series with MMP8 and TMP1 User Guide M212022EN F 8...
Page 116: ......
Page 117: ......
Page 118: ...www vaisala com...