Parameter
Type
Range
Default
Unit
Description
o3_span
uint8
1 ... 255
100
0.01
O
3
sensor span
correction (1 ... 255 =
0.01 ... 2.55)
Table 27 Linear correction parameters - particles
Parameter
Type
Range
Default
Unit
Description
pm1_zero
int16
−10000 ... 10000
0
0.1 μg/m
3
PM
1
offset (zero)
correction
(−10000 ... 10000 =
−1000.0 ... 1000.0)
pm1_span
int16
1 ... 10000
1000
0.001
PM
1
span (gain)
correction (1 ... 10000
= 0.001 ... 10.000)
pm25_zero
int16
−10000 ... 10000
0
0.1 μg/m
3
PM
2.5
offset (zero)
correction
(−10000 ... 10000 =
−1000.0 ... 1000.0)
pm25_span
int16
1 ... 10000
1000
0.001
PM
2.5
span (gain)
correction (1 ... 10000
= 0.001 ... 10.000)
pm10_zero
int16
−10000 ... 10000
0
0.1 μg/m
3
PM
10
offset (zero)
correction
(−10000 ... 10000 =
−1000.0 ... 1000.0)
pm10_span
int16
1 ... 10000
1000
0.001
PM
10
span (gain)
correction (1 ... 10000
= 0.001 ... 10.000)
Table 28 Device identification parameters (read-only)
Parameter
Type
Range
Unit
Description
serial
Text
−
−
AQT530 device serial number in
Vaisala format
Table 29 Base module identification parameters (read-only)
Parameter
Type
Range
Unit
Description
swver
Text
−
−
Base module firmware version
hwver
Text
−
−
Base module hardware version
base_serial
Text
−
−
Base control board with gas cells
boardid
Text
−
−
Base control board without gas
cells
AQT530 Configuration Guide
M212572EN-C
38
Summary of Contents for AQT530
Page 1: ...M212572EN C Configuration Guide Vaisala Air Quality Transmitter AQT530...
Page 62: ......
Page 63: ......
Page 64: ...www vaisala com...