Action
CLI command and
parameter
CLI reference
Modbus register
addresses
Change temperature unit
(°C/°F)
set
tempunit
0x001C
Adjust linear correction
set
•
co_zero
•
co_span
•
no_zero
•
no_span
•
no2_zero
•
no2_span
•
so2_zero
•
so2_span
•
h2s_zero
•
h2s_span
•
o3_zero
•
o3_span
•
pm1_zero
•
pm1_span
•
pm25_zero
•
pm25_span
•
pm10_zero
•
pm10_span
0x0086 ... 0x0097
Set system time
date
Setting date ( date
command) (page 30)
0x0057 ... 0x005C
See
time over Modbus
interface (page 52)
AQT530, base module,
LPC, and HMP
identification (read-only)
show
•
serial
•
sw_ver
•
hw_ver
•
model
•
unit
•
cal_date
•
base_serial
•
lpc_serial
•
hmp_serial
•
hmp_version
lpc
--info
Printing LPC information
( lpc command)
Modbus device
identification:
• Model
• SW version
• Sensor model
• Serial number
• Calibration date
• Hardware version
Modbus registers:
• Calibration date:
0x0013-0x0015
• Base firmware version:
0x004F-0x0053
• HMP firmware version:
0x0054-0x0056
• AQT serial number:
0x00B4-0x00B7
• HMP serial number:
0x00B8-0x00BB
• LPC serial number:
0x00BC-0x00BF
• LPC firmware version:
0x00F4-0x00F8
Chapter 3 – Configuration
13
Summary of Contents for AQT530
Page 1: ...M212572EN C Configuration Guide Vaisala Air Quality Transmitter AQT530...
Page 62: ......
Page 63: ......