Address
RW
Type
Unit
Register
count
Description
0x0087
RW
int16
-
1
O
3
linear correction gain value (1 ... 255,
corresponding gain values 0.01 ... 2.55)
0x0088
RW
int16
-
1
NO
2
linear correction gain value (1 ... 255,
corresponding gain values 0.01 ... 2.55)
0x0089
RW
int16
-
1
SO
2
linear correction gain value (1 ... 255,
corresponding gain values 0.01 ... 2.55)
0x008A
RW
int16
-
1
NO linear correction gain value (1 ... 255,
corresponding gain values 0.01 ... 2.55)
0x008B
RW
int16
-
1
H
2
S linear correction gain value (1 ... 255,
corresponding gain values 0.01 ... 2.55)
0x008C
RW
int16
ppb
1
CO linear correction offset
−10000 ... 10000
0x008D
RW
int16
ppb
1
O
3
linear correction offset
−10000 ... 10000
0x008E
RW
int16
ppb
1
NO
2
linear correction offset
−10000 ... 10000
0x008F
RW
int16
ppb
1
SO
2
linear correction offset
−10000 ... 10000
0x0090
RW
int16
ppb
1
NO linear correction offset
−10000 ... 10000
0x0091
RW
int16
ppb
1
H
2
S linear correction offset
−10000 ... 10000
0x0092
RW
int16
0.001
1
PM
2.5
linear correction gain value
Valid range: 1 ... 10000 (0.001 ... 10.000)
0x0093
RW
int16
0.001
1
PM
10
linear correction gain value
Valid range: 1 ... 10000 (0.001 ... 10.000)
0x0094
RW
int16
0.1 µg/m
3
1
PM
2.5
linear correction offset
Valid range: -10000 ... 10000
(-1000.0 ... 1000.0)
0x0095
RW
int16
0.1 µg/m
3
1
PM
10
linear correction offset
Valid range: -10000 ... 10000
(-1000.0 ... 1000.0)
0x0096
RW
int16
0.001
1
PM
1
linear correction gain value
Valid range: 1 ... 10000 (0.001 ... 10.000)
0x0097
RW
int16
0.1 µg/m
3
1
PM
1
linear correction offset
Valid range: -10000 ... 10000
(-1000.0 ... 1000.0)