Address
RW
Type
Unit
Register
count
Description
0x004B
R
int16
-
1
Status reported by device
• 0 = Unknown. Status information not yet
available, starting up
• 1 = Ok. Operating normally
• 2 = Degraded. Operating but functionality
degraded
• 3 = Faulty
For more information, see status code from
register 0x004C.
0x004C
R
int16
-
1
Status codes
• 0 = No specific status to report
• 1 = LPC malfunction
• 2 = HMP110 malfunction
0x004F
R
uint16
-
1
AQT base module firmware version major
number
0x0050
R
uint16
-
1
AQT base module firmware version minor
number
0x0051
R
uint16
-
1
AQT base module firmware version build
number
0x0052
R
uint32
-
2
AQT base module firmware version hash
number
0x0054
R
uint16
-
1
HMP device firmware version major number
0x0055
R
uint16
-
1
HMP device firmware version minor number
0x0056
R
uint16
-
1
HMP device firmware version revision number
0x0057
RW
uint16
-
1
System time year
• Read = System time
is read into cache
when 0x0057 is
read. Other
registers return
values from cache
• Write = Register
value is written to
cache, and system
time is updated
when 0x005C is
written
0x0058
RW
uint16
-
1
System time month
0x0059
RW
uint16
-
1
System time day
0x005A
RW
uint16
-
1
System time hours
0x005B
RW
uint16
-
1
System time minutes
0x005C
RW
uint16
-
1
System time seconds
0x0065
RW
uint16
-
1
Select the ppb to µg/m
3
conversion type for
registers 0x0066 - 0x006B
• 0 = EU standard (temperature t = +20 °C,
pressure p = 1013.25 hPa) (default)
• 1 = International standard (WHO,
temperature t = +25 °C, pressure
p = 1013.25 hPa)
• 2 = Temperature based dynamic (ambient
temperature from HMP110, pressure
p = 1013.25 hPa)
AQT530 Configuration Guide