Measurement Range Message
Cause
RCD
UB>UL
Contact voltage has exceeded pre-
123
Ω
set limit, measurement
53V UL:50V
interrupted.
RCD
UB>UL
Contact voltage has exceeded pre-
UL:50V
set limit during the socket test.
ZL, IPSC
UB>UL
The contact voltage exceeded the
RCD
---
pre-set limit already during the
UL:50V
socket test. Measurement therefore not carried out
ZL, IPSC
TELARIS 0100 is overheated,
RCD
therefore no further
measurements can be carried out.
ZL, IPSC
Socket incorrectly connected, or
RCD
RCD has tripped.
ZL, IPSC
Insert plug after turning it by
RCD
180°C.
All functions
reference to limit excess within
one measurement variable.
All functions
“
M
”
Minimum one measurement result
has been stored in memory.
All functions
Batteries are virtually empty and
have to be replaced.
Measurement results no longer
comply with the specifications.
All functions
Batteries are empty. No funktion of the Telaris
is blinking
All functions
---
TELARIS 0100 is ready for measurement.
All functions
-
TELARIS 0100 is active (active measurement).
47
Display Indications / Error Messages