background image

UTLDM1

Digital Multimeter

GENERAL MAINTENANCE
Warning    

If the current test leads are worn, replace test leads with identical or 

compatible leads: 1000V 10A. 

Warning    

To avoid personal injury or damage to the meter, DO NOT wet the inner 

parts of the meter. Regularly clean the meter case with damp cloth and a small amount of 

detergent. Do not use abrasives or chemical solvents. 

REPLACE BATTERIES & FUSES
Warning    

To avoid incorrect readings and possible electric shock or personal injury, 

when “ 

 ” appears on the display, replace the battery immediately. Turn off the meter 

and disconnect the test probe from the meter before opening the back cover to replace 

batteries or fuses. Gain access to batteries and fuse by using a screwdriver to loosen the  

battery cover screws on the back of the meter and removing the cover. 

GENERAL SPECIFICATIONS

Operating Category: 600V CAT III, pollution degree: 2.

Operating Elevation (< 2000 m)

Operating Conditions: 0~40°C,  <80% RH (do not use meter when temperature <10°C). 

Storage Conditions: -10~60°C, <70% RH (remove the battery). 

Temperature Coefficient: 1 × Accuracy/°C (<18°C or >28°C).

Maximum Voltage between measurement end and ground: 600V DC or 600V AC RMS. 

Fuse protection: mA Grade: fuse F400mA/600V and 10A Grade: fuse F10A/600V

Sampling rate: about 3 times/second. 

Display: 3 1/2 digits 2,000 count LCD display.

Power supply: 1 x 9V battery.

Instruction Manual

English

P R O F E S S I O N A L   V A L U E

Copyright © 2014 UTL. All Rights Reserved

WARRANTY

 

The UTL Digital Multimeter (UTLDM1) is warranted to be free from defects in materials and 

workmanship for a period of one year from the date of purchase. This warranty does not cover 

fuses, disposable batteries, or damage from drops, neglect, misuse, alteration, contamination, 

or abnormal conditions of operation or handling. Resellers are not authorized to extend any 

other warranty on behalf of UTL. To obtain service during the warranty period, contact your 

nearest UTL service center directly. For full warranty details visit us online www.utltest.net.

DC VOLTAGE 

Range

Resolution Accuracy

2V

1mV

± (0.5% + 2 dgts)

20V

10mV

200V

100mV

600V

1V ± (0.8% + 5 dgts)

Overload protection: 

600V RMS

AC VOLTAGE 

Range

Resolution Accuracy

200V

100mV

± (1.0% + 10 dgts)

600V

1V

Overload protection:  

600V RMS

DC CURRENT

Range

Resolution Accuracy

20mA

0.01mA

± (1.0% + 5 dgts)

200mA

0.1mA

10A

0.01A ± (2.0% + 8 dgts)

mA, μA Overload Protection: 

250V RMS

10A  Overload Protection: 

600V RMS

TEMPERATURE

Range

Resolution Accuracy

-20°C to 1000°C

1°C ± (2.0% + 2 dgts)

-4°F to 1832°F

1°F ± (2.0% + 4 dgts)

RESISTANCE

Range

Resolution Accuracy

200Ω

0.1Ω

± (0.8% + 3 dgts)

2kΩ

20kΩ

10Ω

200kΩ

100Ω

2MΩ

1kΩ

Overload protection: 

250V RMS

DIODE

Range

Test Current Open Test Circuit

0.5 to 0.7V 1.2mA

2.7V

BATTERY TEST

Range

Resolution Test Current

1.5V

0.001V 40mA

9V

0.01V 10mA

12V

0.01V 10mA

CONTINUITY

Overload Protection Open Test Circuit

250V RMS

2.7V

17048  03/14

ELECTRICAL SPECIFICATIONS

1

        

yr.

warranty

IEC61010-1

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

Visit us at www.T

estEquipmentDepot.com

Reviews: