13.0 Technical Data
Display:
3
1/2
digit or 3 digit LCD
Reference-Range:
+17°C to 27°C, max. 70% rel. humidity
Temperature-Range:
0°C to 40°C
Relative humidity:
max. 80%
Height above MSL:
up to 2000m
Batterytype:
6x 1,5V Mignon Typ IEC LR6 (size AA)
Current-consumption:
20mA (Stand-by)
Battery life:
for average use approx. 1 year
Auto-Power-Off:
after approx. 5min
Data memory Interface:
9600 Baud, 1 startbit, 8 data bits, even parity, 1 stoppbit
Pollution degree:
2
Protection Category:
2, IP50
Overvoltage Category:
CAT III / 440V against ground
Dimensions:
235 x 103 x 70 mm
Weight:
650g incl. Batteries
Voltage measurement Vac, Vdc:
Measurement Range Vac:
1 ... 440VAC (Frequency f: 10 ... 99.9 Hz)
Measurement Range Vdc:
1 ... ±440VDC
Resolution:
1 V
Tolerance:
±(3% rdg. ±+3 Digit)
Internal Resistance:
280 k
Ω
Frequency Measurement f:
Measurement Range:
10,0 ... 99,9Hz
Resolution:
0,1Hz
Tolerance:
±(1% rdg. +1 Digit)
Voltage Range:
5 ... 440VAC
Insulation Measurement Rins:
Measurement Range:
0,05 ... 19,99M
Ω
, 20,0 ... 199,9M
Ω
Resolution:
0,01M
Ω
, 0,1M
Ω
Tolerance:
±(5% rdg. 5 Digit)
Nominal output voltage Un:
500V dc
Nominal current In:
≥
1mA at Un
Short-circuit current Ik:
approx.. 3mA
Max. open-circuit voltage:
Un + 50%
Measurement time:
1 - 3s (automatic)
1 - 10s (automatic, at capacitive Load)
Number of Measurements:
1600 (500k
Ω
, 1 measurement - 25s break)
Automatic measurement range selection
EUT automatically discharged after measurement
50
Technical Data
Summary of Contents for Telaris 0100
Page 2: ...2...
Page 28: ...TELARIS 0100 UNITEST Instruction Manual Cat No 9057...
Page 53: ...TELARIS 0100 UNITEST Mode d emploi R f No 9057...
Page 54: ...54...
Page 80: ...80...
Page 81: ...TELARIS 0100 UNITEST Instrucciones para el manejo No 9057...