5.CALIBRATION
28
5-2. Calibration Procedure
Release calibration inhibition lock by switching OFF the Lock Switch on
the rear panel. (Hardware Lock)
Release Setting Value Lock which inhibits the calibration. (Software Lock)
Select Load cell Excitation between 10V to 5V.
Register the Unit of the scale.
Register the position of Decimal point.
Register the maximum capacity of the scale.
If the registered value is exceeded by 9 scale divisions, display shows over
scale "OFL2".
Register the minimum scale division of the scale.
Register the value of load (balance weight) for calibration.
Calibrate the initial zero point.
Place the load (balance weight) on the load cell and calibrate the span
value.
If it is necessary, re-calibrate initial zero point.
Disable changing setting value related to calibration.
Turn on Lock Switch, disable calibration.
LOCK release
LOCK switch OFF
Load cell excitation
Unit
Decimal point
Capacity
Minimum scale division
Balance weight
Zero calibration
Span calibration
Zero calibration
Calibration lock
LOCK switch ON