5.CALIBRATION
34
5-5. Zero Calibration
Register the initial zero point.
- Verify there are no excess loads applied to load cell or scale.
- Check
in green appears on the comparison screen.
Operation
1) Press the
button.
2) Select the "CALIBRATION" tag.
3) Press “ZERO CALIBRATION” tag.
4) Register zero point by pressing
.
5) Zero calibration is completed when the
changes to
and numeral display become to
“0” on the comparison screen.
If a calibration error is displayed, redo zero calibration following the description on the message
screen.
- CAL. ERR. 2:
The initial dead load is above zero adjustment range.
- CAL. ERR. 3:
The initial dead load is negative.
STAB
MODE
OK
CALZ
RUN