14.TROUBLE SHOOTING
99
CAL. ERR. 2 (Calibration error)
Initial Dead Load is above Zero adjustment range.
Remove any excess load from loadcell or scale. If CAL.ERR.2 is still displayed, connect a resistor
EXC and -SIG. loadcell connections. This should shift the Zero point.
Do Zero Calibration again.
- This table is for a 350 ohm loadcell.
When 4 load cells are connected in parallel, the sensitivity decrease to 1/4. Be aware that the
resistance also decreases to 1/4 accordingly.
- The temperature coefficient of the connected resistor directly influences the accuracy of the
indicator. Use a resistor with a temperature coefficient below 50ppm.
Resistor
Strain
Ideal
Approx.
μ-STRAIN
mV/V
875
kΩ
866
kΩ
200
0.1
437
kΩ
442
kΩ
400
0.2
291
kΩ
294
kΩ
600
0.3
219
kΩ
221
kΩ
800
0.4
175
kΩ
174
kΩ
1000
0.5
146
kΩ
147
kΩ
1200
0.6
125
kΩ
124
kΩ
1400
0.7
109
kΩ
110
kΩ
1600
0.8
97
kΩ
97.6
kΩ
1800
0.9
87.3
kΩ
86.6
kΩ
2000
1.0
79.4
kΩ
78.7
kΩ
2200
1.1
72.7
kΩ
73.2
kΩ
2400
1.2
67.1
kΩ
66.5
kΩ
2600
1.3
62.3
kΩ
61.9
kΩ
2800
1.4
58.2
kΩ
57.6
kΩ
3000
1.5
54.5
kΩ
54.9
kΩ
3200
1.6
51.3
kΩ
51.1
kΩ
3400
1.7
48.4
kΩ
48.7
kΩ
3600
1.8
45.9
kΩ
46.4
kΩ
3800
1.9
43.6
kΩ
43.2
kΩ
4000
2.0
41.5
kΩ
41.2
kΩ
4200
2.1
39.6
kΩ
39.2
kΩ
4400
2.2
37.9
kΩ
38.3
kΩ
4600
2.3
36.3
kΩ
36.5
kΩ
4800
2.4
34.8
kΩ
34.8
kΩ
5000
2.5
Connect the resistor
bEXC and -SIG.
+EXC
+SIG
-EXC
-SIG