![Unipulse F701+ Operation Manual Download Page 139](http://html.mh-extra.com/html/unipulse/f701/f701_operation-manual_843485139.webp)
132
8
List of Setting Items
132
Lis
t o
f S
ett
in
g I
te
m
s
Chapter
8
■
Setting mode 4
Name
Initial
value
Setting range
Lock1
(soft)
Lock2
(soft)
LOCK
(SW)
Display
only
Communication
command
1 D/A output mode
00
00 to 21
◎
W50(R/W)
Test mode
0
2: 20mA fixed output
1: 4mA fixed output
0: Weight value linked
Output weight mode
0
1: Net weight
0: Gross weight
2 D/A zero output weight
000.00 0 to 99998
◎
W51(R/W)
3 D/A full scale setting
100.00 1 to 99999
◎
W52(R/W)
4 RS-232C/RS-485 I/F setting
10101 00000 to 61217
Baud rate selection
1
6: Printer mode
(only with RS-232C)
5: 115.2kbps
4: 56800bps
3: 38400bps
2: 19200bps
1: 9600bps
0: 4800bps
Character length
0
1: 8bits
0: 7bits
Parity bit
1
2: Even
1: Odd
0: None
Stop bit
0
1: 2bits
0: 1bit
Terminator
1
(With RS-232C)
7: Communication mode 6
6: Communication mode 5
5: Communication mode 4
4: Communication mode 3
3: Communication mode 2
2: Communication mode 1
1: Communication mode 0/
CR+LF
0: Communication mode 0/
CR
(With RS-485)
2: Modbus-RTU
1: Communication mode 0/
CR+LF
0: Communication mode 0/
CR
5 ID setting
0000 0000 to 9999
◎
6 Transmission delay time
00
00 to 99
◎