0x0223
COMM SAVE DATA
READ NOT AVLB.
Reading saved EEPROM (sensor/slave) data error
Function not supported by slave
0x0224
COMM SAVE DATA
READ RANGE
Reading saved EEPROM (sensor/slave) data error
Too much data was requested from EEPROM
0x0225
COMM SAVE DATA
WRITE NOT OK
Writing saved EEPROM (sensor/slave) data error
Slave responded with “not OK” message.
Possible solution:
More time needed between writing and reading of
EEPROM data subsequently
0x0226
COMM SAVE DATA
WRITE NOT CON.
Writing saved EEPROM (sensor/slave) data error
No slave EEPROM connected at required EEPROM number
0x0227
COMM SAVE DATA
WRITE NOT AVLB.
Writing saved EEPROM (sensor/slave) data error
Function not supported by slave
0x0228
COMM SAVE DATA
WRITE RANGE
Writing saved EEPROM (sensor/slave) data error
Too much data supposed to be written to EEPROM
0x0229
COMM SAVE DATA
EEPROM NOT DEF.
Reading or writing saved EEPROM (sensor/slave) data error
EEPROM is not defined – needs to be defined
0x022A
COMM SAVE DATA
PTR. NOT SET
Reading or writing saved EEPROM (sensor/slave) data error
Pointer to data not set
0x022B
COMM SAVE DATA
BUF. TOO LARGE
Reading or writing saved EEPROM (sensor/slave) data error
Data buffer too large for communication buffer
Solution:
Increase communication buffer or reduce size of data to be
stored on EEPROM
0x022C
COMM SAVE DATA
NO DATA
Reading or writing saved EEPROM (sensor/slave) data error
Size of data not set no data can be stored
0x022D
COMM SAVE DATA
EEPROM NOT FND.
Reading or writing EEPROM (sensor/slave) data error
No EEPROM found – error code of old copy function
9.2.4
Sensor errors
Error code
Display description
Full description
0x0300
SENS CMD BUF FULL
Firmware development error
Command list for a sensor is full
0x0301
SENS UNKNOWN
DATA TYPE
Firmware development error
Unknown data type was loaded with command – values
can not be copied
0x0302
SENS DIVISION BY ZERO
Firmware development error
Values are set in a way so a division by zero could occur
0x0303
SENS UNKNOWN
COMMAND
Firmware development error
Unknown command type was set with command
0x0304
SENS CMD WRONG TYPE
Firmware development error
Type not supported for command list
0x0305
SENS CAL STATUS ZERO IR1
Zero calibration could not be performed on IR channel 1
--- currently not used ---
0x0306
SENS CAL STATUS SPAN IR1
Span calibration could not be performed on IR channel 1
--- currently not used ---
0x0307
SENS CAL STATUS ZERO IR2
Zero calibration could not be performed on IR channel 2
--- currently not used ---
0x0308
SENS CAL STATUS SPAN IR2
Span calibration could not be performed on IR channel 2
--- currently not used ---
0x0309
SENS CAL STATUS ZERO EC1
Zero calibration could not be performed on EC channel 1
--- currently not used ---